Testing Bias Bonds
After bias bonds are done, and before doing any other bonds, one should do the following:
1). Run the I_Leakage vs. V_Bias test. Select "I_Leakage vs. V_Bias" in the "Test" window in the upper left hand corner of the Main Control panel.
End Voltage = 60V.
Step Size = 2V.
1. Total current should never reach 20 uA (current limit of supply).
2. Bias + Guard currents should be equal to Total Current.
3. If Bias and Guard currents are identical, there is probably a short between the bias and guard lines. Check the Upilex for shorted traces and check the wirebonds on the silicon.
2). Run the "I_Leakage vs. Time" test. Select "I_Leakage vs. Time" in the "Test" window in the upper left hand corner of the Main Control panel.
Delta(t) = 0.5 minutes.
Bias Voltage = 60V (if the part did not hit the current limit at 60 V from the previous test (I_Leakage vs. V_Bias)).
People to ask if problems arise: